Pump
Installation and Operating Instructions
The
number one enemy of hydraulic systems is contamination. When replacing
a pump, all filters should be replaced or cleaned and the system
oil checked for cleanliness. Replace dirty or contaminated oil.
To not do so will increase wear on all components of the systemincluding
hydraulic motors!
Pump
rotation is extremely important. Pump damage will occur if the
rotation is in the wrong direction. Rotation is designated by
looking at the pump shaft end. EXAMPLE: With the shaft facing
you and the gear drive on the top, the right side is the HIGH
pressure side with CLOCKWISE ROTATION.
After
filling the reservoir with clean oil and checking all fittings
and connections, be sure the gate valve is open to the pump if
the system has a valve. Many pumps have been ruined because someone
forgot to open the suction side valve to the pump.
NOTE:
Pump inlet oil temperature should not exceed 180 degrees Fahrenheit.
The system also needs a 10 micron or finer return line filter
to keep it clean.
Line size to the pump inlet (suction line) should not be smaller
than the high pressure pump port size.
If
the relief valve adjustment is part of pump, DO NOT adjust it,
as it is preset at the factory. If the relief valve is separate
from the pump, back off the main relief valve until the spring
tension of the adjusting screw is relieved. This procedure will
avoid the possibility of immediate damage to the new pump in the
event that the relief valve setting has been increased beyond
recommended operating pressure prior to removing the old pump.
After
installation the pump must be run at the lowest RPM for 5 minutes.
The pump should run free and not leak, develop heat or excess
noise. Operate for 5 minutes with all valves in the open (neutral)
position and then increase speed and pressure to the desired settings.
A pressure gauge should always be used at the pressure port to
reset the external relief valve to its proper operating pressure.